What the FPIA-3000 deliversSample dispersion and image captureImage extractionParameter
SoftwareData interpretationKey applicationsSpecificationInquiry
FPIA-3000
Specifications
Measurement Range
  High Power Field Mode(2x Secondary lens) Low Power Field Mode(0.5x Secondary lens)
Measurement Range (Standard Unit)Primary lens = 10x 1.5μm - 40μm
(total mag = 20x)
8μm - 160μm
(total mag = 5x)
Measurement Range (High-Mag Unit)Primary lens = 20x 0.8μm - 20μm
(total mag = 40x)
4μm - 80μm
(total mag = 10x)
Measurement Range (Low-Mag Unit)Primary lens = 5x 3.0μm - 80μm
(total mag = 10x)
16μm - 300μm
(total mag = 2.5x)
Sample Volume approx. 5ml
Measurement Time approx. 2 min
Data Storage max. 1,000 samples (Data & Image)
Dimensions and Power Supply
  Dimensions (W x H x D) Weight (Kg) Power Requirement (V) Power Consumption (VA)
Main Unit 900 x 455 x475 60.0 AC 110 300 (50/60Hz)
AC 230
Pneumatic Unit 280 x 400 x 355 20.7 AC 110 230 (50Hz)
280 (60Hz)
AC 230 220 (50Hz)
250 (60Hz)
Required Computer Specifications
OS Windows XP Pro SP2
CPU Pentium4  3.0GHz or higher
Memory

1GB or bigger

HDD 100GB or bigger
Sheath Liquid Compatibility
FPIA-3000 (Standard version) FPIA-3000S (Solvent version)
Aqueous “Particle Sheath Reagent”, Methanol, Ethanol, Isopropyl Alcohol, and Ethylene Glycol Solution (25%) Compatible with most commonly used solvent dispersants such as: Toluene, Acetone, Heptane, Hexane, Methyl Ethyl Ketone, Cyclohexane, Silicone, and Isopar
FPIA-3000 Top Inquiry